Jenway 6300 Benchtop Visible Spectrophotometer; 220 VAC for Parts
Jenway 6300 Benchtop Visible Spectrophotometer; 220 VAC for Parts
The unit powers on, but it gives an error 5 message. Because of this, we must sell it as is for parts or repair.
The manual says the following regarding the error 5 message -
Err 5, No zero order (white) light found during wavelengthcalibration in start up routine. This error is given during thewavelength calibration routine, indicating that even the initialthreshold level was not achieved. Possible causes are; incorrect lamp fitted, cuvette/sample or other obstruction in the light paththrough the sample chamber as well as a possible opticalcomponent failure or misalignment.
Specifications -
- Stray Light<0.5%T
- Source LampTungsten-Halogen
- Bandwidth8.0 nm
- Min Wavelength (nm)320
- Max Wavelength (nm)950
- Wavelength Accuracy±2.0 nm
- Min Photometric - Transmittance (%T)0
- Max Photometric - Transmittance (%T)199.9
- Min Photometric - Absorbance (A)-0.3
- Max Photometric - Absorbance (A)1.999
- Min Photometric - Concentration (C)-300
- Max Photometric - Concentration (C)1999
- DetectorSolid-state
- OutputRS-232; 0 to 1.999 V
- Power (VAC)230
- Power (Hz)50
- DescriptionBenchtop Visible Spectrophotometer; 220 VAC
Regular price
$390.44 USD
Regular price
Sale price
$390.44 USD
Unit price
/
per
Specifications
Product Identifiers
- Brand: Jenway
- MPN: 6300
- eBay Product ID (ePID): 16056264134
Product Key Features
- Type: Spectrophotometer
- Model: Jenway 6300