X-ray Unit H2
Product Basic Information
Product Name:Dental X-ray System
Product Description
1.Power Adapter Input Voltage Uniphase AC 100 ~240V,±10%
Supply Frequency 50/60Hz,±1Hz
Input Power 286W
2.Output Power 25.2W
3.Charge voltage DC 25.2V
4.Charge current 1.0A
5.Nominal value of the focal point 0.8mm
6.Anode angle 19º
7.Available maximum tube voltage at 2mA 60kV
8.Maximum output electric power 0.12kW
9.Maximum loading factor combination 60kV 、2mA 、2s
10.Repeatability of radiation output No more than 0.5
11.Tube voltage (The deviation is within the range of ±10%) 60kV
12.Tube current (The deviation is within the range of ±20%) 2mA
13.Running mode Discontinuous operation, maximum
excitation (open) time : minimum
excitation (close) time = 1:30
14.Total filtration of X-ray tube assembly >2.5mm Al
15.Intrinsic filtration of X-ray tube assembly 0.75mm Al
16.Additional filtration 1.00mm Al
17.Half-value layer ≥1.5mm Al(60kV)
18.Leakage radiation ≤0.25mGy/h(60kV 、2mA 、2s)
19.Adjustment range of the loading time 0.01 ~2.00s,adjustable
20.Distance of the Focal point to skin (The deviation is within 0~5%) 211mm±3mm
21.X-ray field size at the end of beam limiting tube round ,Φ 56mm±3mm
22.Nominal electric power 0.12kW(60kV 、2mA 、0.1s)
23.Available maximum tube current at 60kV 2mA
24.Inverter frequency of the high voltage generator 70kHz
Note: the loading time adjustment range meets the R10 series for the adjustment.
X-ray tube technical specification
1.The X-ray tube model XD10D-0.5 / 70
2.Nominal value of the focal point 0.8mm
3.Target material Tungsten
4.Maximum working tube voltage 70kV
5.Anode angle of the X-ray tube 19º
6.Intrinsic filtration 0.75 mm Al
7.Filament current 2.0 A
8.Filament voltage 2.85 ± 0.5 V
Dental X-ray Sensor R2
Sensor Size: 1.5
Features:
The product is mainly composed of a sensor, an adaptor, a USB connection cable and workstation software. Working together with an X-ray equipment, the Product is intended to make X-ray radiography in dental clinics or veterinary clinics,Sensor size Suitable for adult and child.
Hardware Requirements:
Central Processor: Intel Core i3 or higher
Internal Memory: 4GB or higher
Hard Disk: 80GB or higher
Display: 1280X1024
USB interface: two
Keyboard: One
Mouse: One
Software Requirement
Operating Software:Windows 7 SP1,8,10 32/64bit
Runtime: .Net Framework 4.6 or higher, VC++ 2015
Specification:
Product Name: Image RVG x-ray Sensor
Operating Voltage:5VDC
Input power:500mW
Spatial Resolution:≥7Lp/mm
Low contrast resolution :Able to display holes with 1mm, 1.5mm, 2mm and
2.5mm diameter on the 0.5mm-thick aluminium foil
Image uniformity ≤2%
Effective image area:32*26.5mm
Sensor Size:38.4mm×31.5mm×6mm
Adaptor Size:113.66mm×40mm×15.8mm
Total Cable Length: 3 m (sensor cable + control box cable to PC)